F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
E

Failure to Maintain Functional Call Bell System on 500 Hall

Kadima Rehabilitation & Nursing At GreenvilleGreenville, Pennsylvania Survey Completed on 03-11-2026

Summary

The facility failed to ensure that the call bell system was adequately working on one of six halls (500 hall/Unit 1), as required by its policy that mandates a call bell or alternative device be within reach of each resident in their room, toilet, or bathing area, with staff alerted by visual and audible signals. The installed call system used an oblong digital unit at the nurse's station that displayed two-digit numbers corresponding to room and bathroom call bells and emitted three beeps when the first call was activated. However, the system did not provide ongoing audible alerts, did not illuminate overhead lights down the hall or outside resident rooms, and could only display up to eight active calls at a time. When more than one call bell was activated, subsequent calls displaced earlier numbers on the display without additional sound or visual alerts, and any calls beyond eight were not displayed until earlier calls were cleared. Staff reported that to identify which resident was calling, they had to go to the nurse's station, view the two-digit number on the unit, and then cross-reference a separate list to match the number to a resident room. On the 500 hall, testing and observation revealed that call bell number 47, associated with a specific room, did not display on the digital unit when activated on two consecutive days, a problem confirmed by a CNA and the DON. A resident and family member reported that call bells had not been working adequately for several months, and that the resident had been provided a tap bell by the facility while the daughter purchased a cow bell so staff could hear it. Another resident representative reported that the call bell was not working for that resident. Staff interviews confirmed they had no way to know if more than one call bell was activated unless they physically went to the nurse's station to look at the digital unit, and that the system’s limited display and lack of continuous audible or visible alerts impeded their awareness of active calls. These observations and interviews demonstrated that the call system on the 500 hall was not reliably functioning as intended by facility policy.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0919 citations
Nonfunctioning Call Light and Inaccessible Bell for Dependent Resident
F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
Short Summary

Surveyors found that a cognitively impaired, functionally dependent resident with aphasia did not have a working bedside call light on multiple observations, and the alternative bell was placed out of reach on top of a mini refrigerator. The resident’s care plan and MDS documented extensive ADL assistance needs and fall/safety precautions. The Maintenance Director reported being unaware of the inoperable call light despite an equipment rounding program, while the Administrator described bedside bells as a matter of resident preference rather than a substitute for a nonfunctional call light. A CNA stated that staff are expected to keep call lights within reach and report malfunctions, and facility policy required fully functional, accessible call devices in resident rooms and bathrooms with regular testing, which was not followed in this case.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Call Light System Not Functioning in 500 Hall
D
F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
Short Summary

Call Light System Not Functioning in 500 Hall: A facility failed to ensure the call light system worked in the 500 hall bathroom and bathing area. Repeated observations showed the panel light stayed on while the light above the room did not illuminate, and one room's bed B call light did not light at either the door or the panel. The maintenance log also showed repeated call light issues, and the MDS stated the panel was sometimes reset when the error occurred.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Keep Call Lights Within Reach for Dependent Residents
E
F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
Short Summary

Staff failed to keep call systems within reach for three LTC residents who were incontinent and dependent on staff for ADLs, including residents with arthritis, bipolar disorder, chronic pain, stroke with one-sided impairment, and hemiplegia/hemiparesis. Surveyors observed call lights and call pads placed toward the head of the bed, behind the bed on a light fixture, or hanging on the wall above the bed, all out of the residents’ reach during multiple observations. These practices did not follow the facility’s policy requiring call lights to be within reach and accessible while residents are in bed.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Call Light Not Left Within Reach for Dependent Resident
D
F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
Short Summary

A resident with dementia, anxiety disorder, chronic respiratory failure, and a documented need for substantial assistance with bed mobility was observed with her call light hanging from the bed rail out of her reach. Three pillows were stacked on the side where the call light cord was located, further preventing her from accessing it. An RN confirmed that the call light was not within the resident’s reach, resulting in a cited deficiency related to the call system.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Non-Functional Bathroom Emergency Call System for Cognitively Impaired Resident
E
F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
Short Summary

A resident with Alzheimer’s disease, severe cognitive impairment (low BIMS score), and communication difficulties was care planned for supervision with toileting and partial assistance with bathing, yet was observed ambulating independently to a shared bathroom where the emergency pull-cord system was not functioning. Surveyors found that pulling the bathroom emergency cord did not activate lights or an alert at the nurse’s station, and a CNA was unaware whether the cord signaled at the station. This confirmed that a working emergency call system was not available in the bathroom and bathing area used by the resident.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Maintain Functional Call System for Multiple Residents
D
F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
Short Summary

The facility failed to maintain a functional call system for three residents on one hallway, resulting in non‑working call lights in bedrooms and bathrooms and, in one case, the complete absence of a call light. One resident, care planned to use a call light, instead received a drum she could not effectively use, requiring her to yell or wait for staff checks. Another resident with a traumatic brain injury and convulsions reported having no call light or alternative device and having to walk to the nurses’ station for help. A third resident with diabetes and anxiety also reported a non‑functioning call light and no alternative call system, stating he had to search for staff. The Administrator and a CNA confirmed the south hallway call lights had been inoperative for an extended period, and the acting Maintenance Director acknowledged awareness of the problem and the importance of a working call system.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Know what gets cited — and walk into your next survey with full visibility

We process and analyze inspection reports and Plans of Correction using AI to surface insights and trends — so you can improve care quality and stay ahead of compliance risk before your next survey.

Get ready for your next survey

See what surveyors are citing in your state and spot your risk areas before they do.

Monthly Citation Reports

Have you been cited for this tag?

Save hours drafting a compliant Plan of Correction — AI built on real approved POCs.

Plan of Correction Writer

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release June 24, 2026) and official state health department websites — never guesswork.

Trusted by long-term care providers and associations.

Allegria Senior Living logo
FHCA logo
WeCare Centers logo
Care Rehab logo
An unhandled error has occurred. Reload 🗙