F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
L

Failure to Respond to Pulse Oximetry Alarms for Tracheostomy-Dependent Resident

St Marys Hospital For ChildrenBayside, New York Survey Completed on 03-31-2026

Summary

The deficiency involves the facility’s failure to ensure that a resident requiring respiratory care and continuous pulse oximetry monitoring received services consistent with professional standards of practice and the resident’s care plan. The resident had spastic quadriplegic cerebral palsy, severe hypoxic ischemic encephalopathy, and chronic respiratory failure, was severely cognitively impaired, and was totally dependent on staff for all ADLs. The care plan and physician’s orders required mechanical ventilation with CPAP to tracheostomy collar overnight, humidified trach collar oxygen during the day, and maintenance of oxygen saturation above 92%, with pulse oximeter alarm parameters set to alert below 92%. The resident was equipped with a pulse oximeter linked to the Vocera alert system, which generated alarms at the bedside and on staff mobile devices when oxygen saturation fell outside ordered parameters. On the day of the incident, the resident’s oxygen saturation dropped to 84% at 8:58 AM, triggering an alert to the primary RN via the Patient Safe Solutions/Vocera system, followed by sequential escalation to the buddy RN, the charge RN, and the RT when not acknowledged. The Call Point Detailed Activity Report showed that an alert was sent to the primary RN at 8:58 AM, to the buddy RN at 8:59 AM, and to the charge RN and RT at 9:01 AM. The primary RN pressed “Accepted” on the device at 9:04 AM, and again when the system alerted at 9:17 AM and 9:18 AM, but did not go to the resident’s room to assess the resident and did not document any assessment or intervention. The buddy RN reported not recalling hearing the alert and stated they were administering medications and unaware of the resident’s distress until the rapid response was called. The charge RN acknowledged receiving the alert but did not respond timely, stating they expected the primary or buddy nurse to respond. The RT stated they received the alert but were busy with other residents and expected other staff to respond. From 8:58 AM to 9:23 AM, no assigned nurse or RT responded to the alarms or performed a clinical assessment of the resident, and the alarm cycle continued without intervention. At 9:23 AM, a second alert was triggered when the resident’s oxygen saturation dropped to 52%. An RN who was not assigned to the resident heard an alarm while passing the room, entered, and found the resident in a wheelchair, unresponsive with gray skin. This RN activated a rapid response/Code Blue, assisted in returning the resident to bed, and another RN began chest compressions. EMS was called and arrived at 9:44 AM; a pulse was briefly restored, and the resident was placed on a ventilator and transferred to the hospital, where they were determined to have no brain activity. Life support was later terminated and the resident expired. The facility’s own investigation concluded that nursing and respiratory staff failed to respond to alarms, failed to appropriately acknowledge and review alerts, failed to maintain accessibility to required communication devices, and failed to escalate when they were occupied or unable to respond, resulting in actual harm and Immediate Jeopardy to the resident and placing other monitored residents at risk.

Removal Plan

  • Review camera footage, Patient Safe Solution phone verification notifications, and the pulse oximetry policy.
  • Re-educate involved staff on pulse oximetry alarm response, notification handling, and escalation expectations.
  • Send voice alarm presentation via email to all assistant nurse managers and assistant directors of nursing for review during evening and morning huddles.
  • Ensure Vocera device functionality is reviewed and staff are instructed to keep devices accessible and operational.
  • Have IT/MIS check and confirm monitoring equipment is functioning properly.
  • Implement disciplinary action for staff involved.
  • Discuss and initiate a root cause analysis.
  • Review and revise the pulse oximetry policy.
  • Provide leadership oversight.
  • Implement an audit of alert response times.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0695 citations
Unsecured Storage of Full Oxygen Cylinders on Nursing Unit
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Surveyors observed four full O2 cylinders on one nursing unit stored unsecured directly on the floor under a sign labeled "FULL CYLINDERS" instead of in a secured storage rack. The ADON confirmed the cylinders were full and should not be on the ground. Reference to NFPA 99 showed that freestanding cylinders must be protected from damage and properly chained or supported in a stand or cart. The DON and Maintenance Director both acknowledged that O2 cylinders are required to be stored in a secure rack, should never be on the floor, and that unsecured cylinders on the floor present a safety risk.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Improper Handling and Storage of Oxygen Nasal Cannula
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with severe cognitive impairment and COPD, receiving oxygen therapy via nasal cannula, was observed twice with the cannula lying on the floor beside the bed instead of stored in the bag on the oxygen concentrator as required. A CNA later picked up the cannula from the floor, wiped it with a non-disinfectant incontinent wipe, and reapplied it to the resident, despite having been trained that a cannula found on the floor should be replaced. An LVN, the DON, and the Administrator all confirmed that oxygen cannulas must be stored properly, replaced if found on the floor, and that incontinent wipes are not disinfectants, indicating a failure to follow the facility’s infection prevention and control policy.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Improper Storage of Nebulizer Mask and Respiratory Supplies
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with COPD and dementia, receiving scheduled nebulizer treatments, was found on multiple occasions to have a nebulizer mask stored on top of the machine rather than in a sanitary manner. A CNA and a nurse aide in training confirmed the mask’s placement, and an LPN reported that masks were routinely cleaned, dried, and then stored on top of the machine. The DON later acknowledged that masks should be washed, dried, and placed on a clean surface, and facility policy required oxygen and respiratory supplies to be stored in a plastic bag when not in use.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Missing Current Physician Order for Oxygen
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident was observed receiving O2 via nasal cannula on multiple occasions, but the chart had no current physician order for O2. The resident said she had been told after a recent hospitalization to use O2 for 30 days, but that time had passed and she was still using it because staff told her she needed it. The DON confirmed there was no current O2 order; the last order had already been discontinued.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Oxygen Administered Without Required Physician Order
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with acute respiratory failure with hypoxia, pulmonary hypertension, and type 2 diabetes was observed receiving oxygen at 4.5 L/min via nasal cannula without a corresponding physician order in the clinical record. The DON acknowledged that an order should have been in place before oxygen was initiated. Facility policy on supplemental oxygen via nasal cannula requires administration only under a physician or provider order, in alignment with 410 IAC 16.2-3.1-47(a)(6).

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Provide and Document Respiratory Care
E
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Failure to provide and document respiratory care: A resident with a trach had no documented evidence of respiratory rate, depth, and quality being monitored each shift and as needed, despite oxygen orders and trach care needs. Other residents with CPAP, nebulizer, and oxygen therapy had respiratory equipment left out of required storage, missing CPAP settings and care details in orders and care plans, and MAR entries signed by nursing staff even when respiratory staff reportedly completed the equipment changes.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Know what gets cited — and walk into your next survey with full visibility

We process and analyze inspection reports and Plans of Correction using AI to surface insights and trends — so you can improve care quality and stay ahead of compliance risk before your next survey.

Get ready for your next survey

See what surveyors are citing in your state and spot your risk areas before they do.

Monthly Citation Reports

Have you been cited for this tag?

Save hours drafting a compliant Plan of Correction — AI built on real approved POCs.

Plan of Correction Writer

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release June 24, 2026) and official state health department websites — never guesswork.

Trusted by long-term care providers and associations.

Allegria Senior Living logo
FHCA logo
WeCare Centers logo
Care Rehab logo
An unhandled error has occurred. Reload 🗙