F0880 F880: Provide and implement an infection prevention and control program.
J

Failure to Disinfect Shared Glucometer

Asbury Health And Rehabilitation CenterCharlotte, North Carolina Survey Completed on 10-16-2024

Summary

The facility staff failed to adhere to the manufacturer's instructions for cleaning and disinfecting a shared blood glucose meter between resident uses. This deficiency was observed during a survey when Nurse #1 was seen performing blood glucose checks on two residents without disinfecting the glucometer as per the manufacturer's guidelines. The glucometer was not cleaned with the required EPA-approved disinfectant wipes, which are necessary to prevent the spread of bloodborne infections. Instead, Nurse #1 used an alcohol swab, which is not an acceptable practice according to the facility's policy and the manufacturer's instructions. The facility's policy, revised in May 2024, clearly outlines the procedure for disinfecting glucometers, which includes using two disinfectant wipes to clean and disinfect the device thoroughly, followed by a two-minute air-dry time. However, during the observation, Nurse #1 did not follow these steps and admitted to forgetting the procedure due to nervousness. The nurse also lacked knowledge of the required wet and dry times for the disinfectant wipes, indicating a gap in training and adherence to the facility's infection control protocols. Interviews with the Infection Preventionist and the Director of Nursing revealed that the facility had provided education on glucometer disinfection, but Nurse #1 had not received the recent training. The Infection Preventionist confirmed that the facility did not have dedicated glucometers for each resident, relying instead on staff adherence to disinfection protocols. The Director of Nursing expressed surprise at the non-compliance, as the staff had been trained, and the process had not been an issue previously. The deficiency was identified as an immediate jeopardy situation, highlighting the potential risk of spreading bloodborne pathogens among residents.

Removal Plan

  • Identify those recipients who have suffered, or are likely to suffer, a serious adverse outcome as a result of the noncompliance.
  • All residents residing in the building that receive blood glucose monitoring at the time of the observation of non-compliance were identified, especially those that resided in the same household where the non-compliance occurred.
  • All residents residing in household two that could have been affected by the deficient practice were seen by the medical provider, with orders received as necessary by the practitioner's assessment.
  • All glucometers that are presently in the clinical spaces in the building were disinfected, per policy and manufacturer's recommendations.
  • All diagnoses of residents in the building were reviewed to ensure that no one currently has an active diagnosis of a bloodborne pathogen.
  • The policy and procedure for glucometer disinfection was reviewed and compared to manufacturer recommendations.
  • The nurse found to be non-compliant with the glucometer disinfection process was re-educated with return demonstration, as were all nurses in the building at the time of the observation of non-compliance.
  • All nursing staff that do (or could) perform glucose monitoring will be in-serviced on the glucometer disinfection process before being allowed to work.
  • All staff members will also have a skills validation performed to ensure that they can perform the disinfection appropriately.
  • Any staff that do not receive the education and skills validation will not be allowed to work until they are compliant with the educational training.
  • The County Communicable Disease branch was notified of the infection control breach.
  • Communication was also provided to the residents affected by the deficient practice and/or their responsible parties.

Penalty

Inspection fine: $59,040
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0880 citations
Failure to Follow Enhanced Barrier Precautions During Wound Care
D
F0880 F880: Provide and implement an infection prevention and control program.
Short Summary

A resident with a chronic heel wound with drainage, classified as high risk under the facility’s Enhanced Barrier Precautions (EBP) policy, received wound care from a Wound Nurse and a NA who wore masks and gloves but did not don gowns during multiple high-contact wound care activities on both lower extremities. The facility’s EBP policy requires both gloves and gowns for high-contact care, including wound care, for residents with chronic wounds. At the time of care, there was no EBP sign on the door and no PPE caddie or supplies outside the room. In subsequent interviews, the Wound Nurse and NA reported they did not wear gowns because there was no sign on the door and the nurse was not wearing one, while the IP and DON stated they would have expected gown use and confirmed that wound care is considered a high-contact activity under the policy.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Implement and Follow Enhanced Barrier Precautions During Wound Care
D
F0880 F880: Provide and implement an infection prevention and control program.
Short Summary

Staff failed to consistently implement and follow Enhanced Barrier Precautions (EBP) during wound care for two residents. For a resident with an indwelling urinary catheter and an EBP order, an RN and a CNA removed their gowns after catheter care and performed a heel and toe dressing change wearing only gloves, despite a door sign requiring gown and gloves for wound care and other high-contact care. For another resident with multiple open leg wounds and active wound care orders, an RN and a nurse aide performed dressing changes with gloves only, without gowns, and there was no EBP signage or order in place. Interviews with nursing staff, the IP, and the DON revealed inconsistent understanding and application of the facility’s EBP policy, which requires gown and gloves for high-contact care activities, including wound care and device care, for residents with chronic wounds or indwelling devices.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Incomplete COVID Surveillance and Return-to-Work Tracking
F
F0880 F880: Provide and implement an infection prevention and control program.
Short Summary

The facility failed to fully document infection surveillance and RTW decisions during a COVID outbreak. Multiple staff members reported symptoms such as sore throat, headache, congestion, diarrhea, vomiting, fever, and cough, but the employee illness logs were incomplete and left the RTW date blank, with no indication they were tested for COVID or cleared per CDC guidance. At the same time, multiple residents were diagnosed with COVID and others had GI symptoms with unknown testing status. The IP said she worked infection control only a few hours per week and had not thoroughly reviewed the logs for trends, while the DON had not been reviewing the surveillance logs.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Cross Contamination During Dressing Change and Infection Control Program Deficiencies
E
F0880 F880: Provide and implement an infection prevention and control program.
Short Summary

Cross contamination occurred during a dressing change when an LPN placed a resident’s foot directly on the wheelchair seat without a barrier and did not clean the bedside table after the procedure. The facility also lacked infection surveillance documentation for several months, and its Legionella water management plan was incomplete, with no mapping of high-risk areas, no temperature logs, and no documented preventive measures for unused areas.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Infection Control Failures During Resident Care
E
F0880 F880: Provide and implement an infection prevention and control program.
Short Summary

Infection Control Failures During Resident Care: Staff did not follow PPE, hand hygiene, and equipment-cleaning practices during care for several residents. An RN failed to clean a glucometer and basket after blood sugar checks, a CNA and a Central Supply staff member entered rooms with enhanced barrier precautions without PPE, and an LVN did not clean the glucometer or insulin vial, and did not properly perform hand hygiene during insulin administration and after emptying a urinal. Residents involved had significant cognitive impairment, diabetes, wounds, and other serious diagnoses.

Inspection fine: $27,378
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Infection Control Lapses in Laundry Services and Policy Review
F
F0880 F880: Provide and implement an infection prevention and control program.
Short Summary

Infection control failed during laundry services when staff reported using the same personal T-shirt for handling dirty laundry and then hanging clean laundry, while using disposable gowns only for laundry from a resident with an infection. The DON also acknowledged that the Infection Prevention Program policy was overdue for annual review, and the policy showed no indication of an annual review.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across North Carolina

Get a heads-up on the newest immediate-jeopardy (J–L) citations in North Carolina — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release August 26, 2026) and official state health department websites — never guesswork.

In your survey window? See what surveyors are citing.

The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.

Get the Survey-Prep Report
An unhandled error has occurred. Reload 🗙

Connection lost — reconnecting… We couldn't reconnect automatically. Please reload the page to continue.