F0908 F908: Keep all essential equipment working safely.
E

Deficiencies in Kitchen and Pantry Equipment Maintenance

Aviata At TallahasseeTallahassee, Florida Survey Completed on 06-20-2024

Summary

The facility failed to maintain food service-related equipment in safe operating condition, as observed during a survey. In the kitchen, a significant amount of water was found on the floor due to a leaking sprayer nozzle and a pipe under the sink. Additionally, there was a buildup of food particles on the clean side of the dishwasher, and dishes were not properly cleaned, as they contained water, soap bubbles, and food particles. The Kitchen Manager acknowledged these issues but did not provide a satisfactory explanation for the improper cleaning process. The facility's Administrator was aware of the dishwasher concern but failed to address the leaking pipe under the sink. The walk-in refrigerator and freezer also presented issues, with water on the floor, condensation on containers, a broken inner door handle, and a broken thermometer. The temperature in the refrigerator was found to be 60 degrees F, which is above the required 41 degrees F for perishable foods. The freezer had a significant buildup of ice, and the temperature was 10 degrees F. The Kitchen Manager admitted that the refrigerator had trouble maintaining temperature during warm weather and that these issues had persisted for over a year without resolution from the administration. In the pantry rooms, one ice machine was unplugged, and another was leaking water, causing a wet floor. The refrigerators in these pantries had a large buildup of ice, indicating a lack of maintenance. The interim Plant Operations Manager was unaware of these issues, and no work orders or maintenance logs were provided to address these concerns. The facility's policies on warewashing, food storage, equipment maintenance, and ice machine sanitation were not followed, leading to these deficiencies.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0908 citations
Failure to Maintain AC Preventative Maintenance Schedule and Critical Component Testing
E
F0908 F908: Keep all essential equipment working safely.
Short Summary

Surveyors found that the facility did not maintain a documented maintenance schedule for its two AC units and relied only on undocumented daily visual checks by maintenance staff. One AC unit was not working while a belt was being changed, and another had been nonfunctional previously. An AC technician reported that the units required monthly PM, including filter changes and testing of water valves and pneumatic controls, but these tasks were not part of the facility’s PM program. Review of the facility’s maintenance policy showed that the Maintenance Director was required to develop and maintain maintenance schedules for building systems, which was not done, creating the potential for residents in general to lack a comfortable environment.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Frayed bed remotes and nonfunctioning call light
E
F0908 F908: Keep all essential equipment working safely.
Short Summary

Frayed and exposed wires were observed on bed remote controls in the rooms of three residents, including residents with impaired cognition, mobility dependence, and diagnoses such as paraplegia, muscle weakness, and depression. Staff stated the exposed wiring should not have been present because of the risk of electrocution, and one bed remote was also not working when used for care. In a separate room, a resident’s call light did not activate the light outside the door or ring at the nurse’s station, and the resident reported being unable to get help when calling for assistance.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Walk-In Freezer Not Maintained
F
F0908 F908: Keep all essential equipment working safely.
Short Summary

Walk-In Freezer Not Maintained: The facility failed to keep the walk-in freezer free of water drippings and ice build-up. During kitchen observation, the freezer ceiling had numerous frozen water drops and the floor had three frozen areas about 12 inches in diameter. The CD said the condition had been present for a couple of weeks and that the frozen water on the floor was a safety hazard. The CD could not find a work order, and the DM said he was not aware of the current build-up.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Maintain Crash Cart Medications and Equipment in Safe Operating Condition
E
F0908 F908: Keep all essential equipment working safely.
Short Summary

Surveyors found that the crash cart contained multiple expired or out-of-date emergency supplies, including a suction machine overdue for inspection, expired iodine packets, aspirin, a biohazard spill kit, airway tubes, suction components, small bore extension kits, a central line dressing kit, and a heat pack. The DON confirmed the items were expired but reported believing the dates were manufacturing dates and stated that monthly checks of the crash cart were performed using a checklist that did not record expiration dates. Review of facility documentation showed completed checklists with all items marked as present but no tracking of expirations, and an office manager confirmed there was no active crash cart policy in place, despite an undated written policy stating that crash carts would be kept in a constant state of readiness and that expiration dates would be routinely monitored.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Unsafe Shower Chairs with Ineffective Brakes Used During Resident Transfers
D
F0908 F908: Keep all essential equipment working safely.
Short Summary

The facility failed to ensure shower chairs functioned safely, as multiple plastic-framed chairs with plastic casters could slide and roll on tile floors even when wheel brakes were fully engaged. A cognitively intact resident who required partial to moderate assistance for transfers fell when a shower chair moved backward and a wheel came out during a transfer to a wheelchair, despite staff reporting that the brakes were locked. CNAs, an LPN, the DON, and maintenance staff observed and demonstrated that several shower chairs could be easily moved or rolled with brakes applied, and one bariatric chair had locks on only two rear wheels, allowing the front to swing side to side. Staff reported they did not rely on the brakes and instead physically held the chairs during use, and the facility could not provide manufacturer instructions for the shower chairs.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Maintain Functional Kitchen Stove/Oven for Resident Meal Service
E
F0908 F908: Keep all essential equipment working safely.
Short Summary

The facility failed to keep the main kitchen stove/oven in working order, resulting in altered meal preparation for residents over an extended period. During a lunchtime observation, the stove/oven was found nonfunctional and staff were serving cold ham and cheese sandwiches instead of hot meals. The cook stated the stove/oven had been out of service for over 2 months and that the menu had been changed for more than a month, causing resident dissatisfaction. The Dietary Manager and the Nursing Home Administrator both confirmed that the stove/oven had been down for over a month and described unsuccessful attempts to replace it due to incompatible gas and electrical hookups.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across Florida

Get a heads-up on the newest immediate-jeopardy (J–L) citations in Florida — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release August 26, 2026) and official state health department websites — never guesswork.

In your survey window? See what surveyors are citing.

The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.

Get the Survey-Prep Report
An unhandled error has occurred. Reload 🗙

Connection lost — reconnecting… We couldn't reconnect automatically. Please reload the page to continue.